FR2000+ Series Clamp Earth Resistance Tester
- Resistor Resolution:0.001Ω
- Current resolution:0.01mA
- Measurement Method:mutual inductance
- Data Storage:300 groups, "MEM" storage indication, displaying "FULL" symbol means the storage is full
- Measure Time:2 times/sec
Model | Resistance Range | Current Range |
FR2000A+ | 0.01Ω-200Ω | — |
FR2000+ | 0.01Ω- 500Ω | — |
FR2000C+ | 0.01Ω-1200Ω | 0.00mA-20A |
FR2000B+Explosion-proof | 0.01Ω-1200Ω |
Function | Ground resistance test, loop resistance test, leakage current test |
Ambient Temperature and Humidity | 23℃±5℃, below 75%rh |
Power Supply | DC 6V (4 AA dry batteries) |
Measurement Method | mutual inductance |
Resistor Resolution | 0.001Ω |
*Current resolution | 0.01mA |
Jaw size | 55mm×32mm |
*Ω + A sync display | Ω + A sync display |
Ground voltage measurement function | None |
Display Mode | 4-digit LCD display, black screen design |
LCD size | 46mm×29mm |
Meter size | Length, width and height:285mm×85mm×58mm |
Measure Time | 2 times/sec |
Data Storage | 300 groups, “MEM” storage indication, displaying “FULL” symbol means the storage is full |
Data Access | “MR” symbol indication when viewing data |
Overflow Display | “OL” symbol indication when over range overflow |
Interference Test | Automatically identify interference signals, and the “NOISE” symbol indicates when the interference current is large |
Alarm Function | When the measured value exceeds the alarm setting value, an alarm prompt will be issued |
Battery Voltage | Real-time display of battery power, reminding to charge in time when the battery voltage is low |
Automatic shut-down | Automatic shutdown after 5 minutes of startup |
Power consumption | 50mA Max |
Quality | Meter: 1180g (including battery) |
Working temperature and humidity | -10℃~40℃; below 80%rh |
Storage temperature and humidity | -20℃~60℃; below 70%rh |
Insulation resistance | Above 20MΩ (500V between circuit and case) |
Pressure resistance | AC 3700V/rms (between circuit and case) |
External Magnetic Field | <40A/m |
External Electric Field | <1V/m |
Suitable for safety regulations | IEC61010-1 (CAT III 300V, CAT IV 150V, pollution degree 2); IEC61010-031; IEC61557-1 (grounding resistance); |